Answer:
The Mangal MS9000C utilizes an electrochemical fuel cell sensor providing a measurement display range of 0 to 0.4% BAC (0 to 4.0‰ BAC or 0 to 2.0 mg/L) with a standard deviation precision of ±0.025 mg/L or ±5%. It features a 4.5-inch IPS LCD touchscreen, automated flow detection, an operating range of -20°C to +50°C, and is powered by a rechargeable 3.7 V Li-ion battery (they are not included).
The Mangal MS9000C is a law-enforcement-grade breath alcohol analyzer designed around an electrochemical fuel cell sensor. The sensor mechanism consists of a porous disk with a platinum black layer on both faces, releasing electrons in exact proportion to ethanol breakdown for high selectivity and repeatability.
Key technical specifications include:
- Sensor Technology: Electrochemical / Fuel Cell Sensor with automatic chamber and sensor warmup
- Measurement Units: Configurable between mg/L, mg/100mL, % BAC, and ‰ BAC
- Measurement Range: 0 to 0.4% BAC, 0 to 4.0‰ BAC, or 0 to 2.0 mg/L (or higher)
- Measurement Precision: Standard deviation ±0.025 mg/L or ±5% of reading
- Sensor Stability & Drift: Typically under 0.8% of measurement value per month
- Display & Input: 4.5-inch IPS LCD screen with touch panel input
- Sampling Verification: Automated breath flow detection monitoring
- Operating Environment: -20°C to +50°C (optimal testing operation between 5°C and +50°C)
- Storage Conditions: -20°C to +55°C at 30% to 90% RH
- Power Supply: Rechargeable 3.7 V Li-ion battery (they are not included)
- Instrument Weight: 1.2 kg